IDP10-T/IGP10-T Calibrate at Known Tank Level
X = Actual value to be displayed
Y = Original M1 LRV value
- Press the NEXT button and the word RERANGE will be displayed.
- Press ENTER and M1URV will be displayed
- Press NEXT and M1LRV will be displayed
- Make note of this value as the “Y” value
- Press ENTER and left hand most digit will blink. Press NEXT until this digit is equal to the most significant digit of the “X” value
- Press ENTER and the next digit will flash
- Press NEXT until this digit matches the value of the next most significant digit of the X value. Repeat procedure for the remaining digits until the entire X value has been entered
- Press ENTER and Place DP will be displayed
- Press NEXT until the decimal point is in the correct location
- Press ENTER and CANCEL will be displayed
- Press NEXT and SAVE will be displayed
- Press ENTER and device will be returned to Measurement Display
- Press the NEXT button 3 times and CALIB will be displayed
- Press ENTER and display will indicate CAL AT0
- Press next and display will indicate CAL LRV
- Press the ENTER button and display will indicate CAL WAIT and then display LRVdone
- Press NEXT 6 times and SAVE will be displayed
- Press Enter
- Press the NEXT button and the word RERANGE will be displayed.
- Press ENTER and M1URV will be displayed
- Press NEXT and M1LRV will be displayed
- Press ENTER and left hand most digit will blink. Press NEXT until this digit is equal to the most significant digit of the “Y” value
- Press ENTER and the next digit will flash
- Press NEXT until this digit matches the value of the next most significant digit of the Y value. Repeat procedure for the remaining digits until the entire Y value has been entered
- Press ENTER and Place DP will be displayed
- Press NEXT until the decimal point is in the correct location
- Press ENTER and CANCEL will be displayed
- Press NEXT twiceand SAVE will be displayed
- Press ENTER and device will be returned to Measurement Display
- Complete