How to Update the Application for the Vigilohm IMDLRDH

Note: Implementing a new application in your IMDLRDH will erase your previous configuration.

To update the application program of your IMDLRDH you need to:

Have a formatted USB Pen Drive in FAT32 format.

Unzip the content of application zip file and place the extracted files on the root of your USB Key.

Image showing the extracted application file in root of the USB Drive

Switch off the IMDLRDH

Plug the USB Key in the USB port of the IMDLRDH

Image showing USB port of IMLRDH where the USB Drive needs to be inserted

Power on the IMDLRDH, wait until the installation confirmation screen appears.

Image showing the confirmation to install the application file

Press Yes to Start the installation.

Device will restart and installation will start automatically.

Image showing the installation progress

Wait for the program to be installed completely in the IMDLRDH

Once installation is complete this screen will appear.

Image showing the installation is complete

Remove the USB Pen Drive and click restart.

After these steps you need to perform the commissioning of your IMDLRDH.

Schneider Electric Казахстан

Узнайте больше
Линейка:
Узнайте больше
Линейка:
  • Документация по продукту
  • Загрузка ПО
  • Система подбора продукции
  • Замена продукта
  • Центр справки и связи
  • Найти наши офисы
  • Получить коммерческое предложение
  • Где купить
  • Сообщество Schneider Electric
  • Карьера
  • Сведения о компании
  • Сообщить о нарушении
  • Доступность
  • Отдел новостей
  • Инвесторы
  • EcoStruxure
  • Поиск вакансий
  • Блог
  • Политика конфиденциальности
  • Уведомление об использовании файлов cookie
  • Change your cookie settings