Issue:
What does error code 8.8 mean on the Iso-Gard™ Series 6 IG6 Line Isolation Monitor (LIM)?
Product Line:
Medical Products
Environment:
IG6 LIM
Cause:
Under normal operating conditions, this error appears when the LIM battery is low; however, at first use/initial energization, this same error code 8.8 indicates that the installer/LIM operator needs to reset the clock.
Resolution:
Please see video below as well as the attachment for the process of resetting the clock.
What does error code 8.8 mean on the Iso-Gard™ Series 6 IG6 Line Isolation Monitor (LIM)?
Product Line:
Medical Products
Environment:
IG6 LIM
Cause:
Under normal operating conditions, this error appears when the LIM battery is low; however, at first use/initial energization, this same error code 8.8 indicates that the installer/LIM operator needs to reset the clock.
Resolution:
Please see video below as well as the attachment for the process of resetting the clock.