What does error code 8.1 on IGD series D LIM or Line isolation monitor mean?
Issue:
What does error code 8.1 on IGD series D LIM or Line isolation monitor mean?

Product Line:
LIM -  Line isolation Monitor

Environment:
IGD series D LIM

Cause:
Error code 8.1

Resolution:
LIM displays Er (error code) 8.1 either continuously or intermittently. The LIM may have been in operation for some time or may display this problem on start up.

Possible cause: One of the field connected lead wires from the LIM terminal board to the
remote indicator alarm (Cat. No. IA-NC) may have a short circuit to ground.

Action: Disconnect power to the LIM. Remove the field connected remote indicator lead
wires at the LIM terminal board. Re-energize the LIM; if the problem is corrected
examine the lead wires to the remote indicator for a possible short circuit to ground
using an ohmmeter. The impedance between any of the three remote indicator lead
wires and ground should be very high (greater than 1 Mohm). Repair the short circuit
to ground and reconnect the lead wires to the remote indicator. Re-energize the LIM to
verify repair of the short circuit.

Note: The use of solid conductors is not recommended for the field connected remote
lead wires.

Schneider Electric USA

Users group

Discuss this topic with experts

Visit our community and get advice from experts and peers on this topic and more
  • Products Documentation
  • Software Downloads
  • Product Selector
  • Product Substitution and Replacement
  • Help and Contact Center
  • Where to buy
  • Get a Quote
  • Find our Offices
  • Schneider Electric Community
  • Careers
  • Company Profile
  • Report a misconduct
  • Accessibility
  • Newsroom
  • Investors
  • EcoStruxure
  • Job Search
  • Blog
  • Tech Support: 877-342-5173
  • Privacy Policy
  • Cookie Notice
  • Terms of use
  • Change your cookie settings