Our Brands

Schneider Electric USA Website

Welcome to our website.
		
How can we help you today?
What does error 9.31 mean for an IG6M Line isolation Monitor
Issue:
Why does the IG6M LIM keep alarming every hour with a 9.31 error displayed?

Product Line:
Medical Products

Environment:
Troubleshooting

Cause:
Alarms

Resolution:
The error 9.31 points to the IG6 LIM having a firmware issue with it’s internal memory and every hour when it does it’s self diagnostic test, it tests it's memory causing the alarm to appear.
The only recommendation is to replace the IG6M LIM with another one

Schneider Electric USA

Users group

Discuss this topic with experts

Visit our community and get advice from experts and peers on this topic and more
  • Products Documentation
  • Software Downloads
  • Product Selector
  • Product Substitution and Replacement
  • Help and Contact Center
  • Where to buy
  • Get a Quote
  • Find our Offices
  • Schneider Electric Community
  • Careers
  • Company Profile
  • Report a misconduct
  • Accessibility
  • Newsroom
  • Investors
  • EcoStruxure
  • Job Search
  • Blog
  • Tech Support: 877-342-5173
  • Privacy Policy
  • Cookie Notice
  • Terms of use
  • Change your cookie settings