How do you clear error code 8.11 on the IG6 LIM (line isolation monitor)?
Issue:
IG6 - What does error code 8.11 mean?
Product Line:
Medical products
Environment:
IG6, IG6M - Line Isolation Monitor (LIM)
Cause:
Error code 8.11
Resolution:
Error code 8.11 indicates there is an issue with the control board and the unit must be replaced. This could happen any time from shipping to installation to operation.
IG6 - What does error code 8.11 mean?
Product Line:
Medical products
Environment:
IG6, IG6M - Line Isolation Monitor (LIM)
Cause:
Error code 8.11
Resolution:
Error code 8.11 indicates there is an issue with the control board and the unit must be replaced. This could happen any time from shipping to installation to operation.