What does error 9.31 mean for an IG6M Line isolation Monitor
Issue:
Why does the IG6M LIM keep alarming every hour with a 9.31 error displayed?
Product Line:
Medical Products
Environment:
Troubleshooting
Cause:
Alarms
Resolution:
The error 9.31 points to the IG6 LIM having a firmware issue with it’s internal memory and every hour when it does it’s self diagnostic test, it tests it's memory causing the alarm to appear.
The only recommendation is to replace the IG6M LIM with another one
Why does the IG6M LIM keep alarming every hour with a 9.31 error displayed?
Product Line:
Medical Products
Environment:
Troubleshooting
Cause:
Alarms
Resolution:
The error 9.31 points to the IG6 LIM having a firmware issue with it’s internal memory and every hour when it does it’s self diagnostic test, it tests it's memory causing the alarm to appear.
The only recommendation is to replace the IG6M LIM with another one