Procedure for complete uninstall and reinstall of ION Setup

Issue
A user may want to completely uninstall and reinstall ION Setup.

Product Line
ION Setup (All versions)

Environment
Windows Operating System

Cause

User couldn't uninstall ION Setup software from the PC/ Laptop.

Resolution
*Warning: This article contains information about editing the registry. Improper changes to the registry can permanently damage the operating system. Always backup the registry before making any changes.*

Under normal circumstances, ION Setup can be upgraded by running the installer for the new version on a PC on which ION Setup is already installed.

If a user wishes to COMPLETELY uninstall the previous version of ION Setup and start from scratch with a new version, the following procedure should be followed:

1. Uninstall ION Setup through the Control Panel "Programs and Features."
2. Remove (or rename) all traces of ION Setup in the registry.

(HKEY_LOCAL_MACHINE\SOFTWARE\Wow6432Node\Schneider Electric\ION Setup)

regedit
3. Remove (or rename) all traces of ION Setup on the hard disk file system.

(C:\Program Files (x86)\Schneider Electric\ION Setup and C:\ProgramData\Schneider Electric\ION Setup)


4. Install the new version of ION Setup.

Schneider Electric Україна

Досліджуйте більше
Серія:
Досліджуйте більше
Серія:
  • Документація Продукції
  • Завантаження ПЗ
  • Інструмент вибору продукту
  • Заміна та заміщення продукту
  • Контактний центр із надання довідки
  • Знайти наші представництва
  • Де купити
  • Спільнота Schneider Electric
  • Запит цінової пропозиції
  • Вакансії
  • Профіль компанії
  • Повідомити про проступок
  • Спеціальні можливості
  • Новини
  • Інвестори
  • EcoStruxure
  • Вакансії Пошук
  • Блог
  • Загальні умови ведення бізнесу
  • Політика конфіденційності
  • Сповіщення про файли cookie
  • Умови користування
  • Change your cookie settings